Complications from prophylactic replacement of cardiac implantable electronic device generators in response to United States Food and Drug Administration recall: A systematic review and meta-analysis

美国食品药品监督管理局召回后预防性更换心脏植入式电子设备发生器的并发症:系统评价和荟萃分析

阅读:1

Abstract

BACKGROUND: The number of cardiac implantable electronic device (CIED) recalls and advisories has increased over the past 3 decades, yet no consensus exists on how to best manage patients with these CIEDs, partially because rates of complications from prophylactic replacement are unknown. OBJECTIVE: The purpose of this study was to establish rates of complications when recalled CIED generators are replaced prophylactically. METHODS: We searched MEDLINE and the Cochrane Controlled Trials Register for reports of prophylactic replacement of recalled CIED generators. Studies with <20 subjects were excluded. We then conducted a meta-analysis of qualifying studies to determine the rates of combined major complications, mortality, and reoperation. RESULTS: We identified 7 citations that met our inclusion criteria and reported ≥1 end-points of interest. Four were single center, and 3 were multicenter. Six studies collected data retrospectively (n = 1213) and 1 prospectively (n = 222). Using a random effects model to combine data from all included studies, the rate of major complications was 2.5% (95% confidence interval [CI] 1.0%-4.5%). Combining data from 6 studies reporting mortality and reoperation, the rates were 0.5% (95% CI 0.1%-0.9%) and 2.5% (95% CI 0.8%-4.5%), respectively. CONCLUSION: Prophylactic replacement of recalled CIED generators is associated with a low mortality rate but nontrivial rates of other major complications similar to those reported when CIED generators are replaced for other reasons. Thus, when considering replacing a recalled CIED generator, known risks of elective generator replacement likely apply and can be weighed against risks associated with device failure.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。