Evaluation of ablation catheter technology: Comparison between thigh preparation model and an in vivo beating heart

消融导管技术评价:大腿预备模型与体内跳动心脏的比较

阅读:2

Abstract

BACKGROUND: An in vivo animal thigh model is the standard technique for evaluation of ablation catheter technologies, including efficacy and safety of ablation. However, the biophysics of ablation in a thigh model may not be similar to a beating heart. OBJECTIVE: The purpose of this study was to compare efficacy and safety of ablation between a thigh preparation model and a beating heart. METHODS: In 7 swine, radiofrequency ablation using a 3.5-mm open irrigated catheter (ThermoCool Smart Touch) was performed sequentially in a thigh muscle and in vivo beating ventricles. Ablation was performed at low (30 W for 40 s) and high (40 W for 60 s) energy settings and at similar contact force. Ablation lesions were scanned in high resolution and measured using electronic calipers. RESULTS: A total of 152 radiofrequency ablation lesions were measured (86 thigh and 66 heart). At low energy, lesion width was greater in the thigh model (12.19 ± 1.8 mm vs 8.99 ± 2.1 mm; P <.001), whereas lesion depth was similar between the thigh and heart (5.71 ± 0.8 mm vs 5.95 ± 1.3 mm, respectively; P = .18). The planar cross-sectional lesion area was greater in the thigh model (thigh 54.8 ± 10.8 mm(2) vs heart 43.1 ± 16.1 mm(2); P <.001). At the high-energy setting, lesion depth, width, and area were all greater in the thigh model (thigh 91.5 ± 16.8 mm(2) vs heart 56.0 ± 15.5 mm(2); P <.001). The incidence of steam pop and char formation was similar between the models. CONCLUSION: The thigh preparation model is a reasonable technique for evaluation of ablation catheter technology; however it often results in overestimation of lesion size, especially at higher energy settings.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。