Periodic Comparability Verification and Within-Laboratory Harmonization of Clinical Chemistry Laboratory Results at a Large Healthcare Center With Multiple Instruments

大型医疗中心多台仪器临床化学实验室结果的定期可比性验证和实验室内部协调

阅读:1

Abstract

BACKGROUND: Results from laboratories using multiple instruments should be standardized or harmonized and comparability-verified for consistent quality control. We developed a simple frequent comparability verification methodology applicable to large healthcare centers using multiple clinical chemistry instruments from different manufacturers. METHODS: Comparability of five clinical chemistry instruments (Beckman Coulter AU5800, Abbott Architect Ci16000, two Siemens Vista 1500, and Ortho Vitros 5600) was evaluated from 2015 to 2019 for 12 clinical chemistry measurements. Pooled residual patient samples were used for weekly verifications. Results from any instrument exceeding the allowable verification range versus the results from the comparative instrument (AU5800) were reported to clinicians after being multiplied by conversion factors that were determined via a linear regression equation obtained from simplified comparison. RESULTS: Over the five-year study period, 432 weekly inter-instrument comparability verification results were obtained. Approximately 58% of results were converted due to non-comparable verification. Expected average absolute percent bias and percentage of non-comparable results for non-converted and converted results after conversion action were much lower than those for data measured before conversion action. The inter-instrument CV for both non-converted and converted results after conversion action was much lower than that for measured data before conversion action for all analytes. CONCLUSIONS: We maintained within-laboratory comparability of clinical chemistry tests from multiple instruments for five years using frequent low-labor periodic comparability verification methods from pooled residual sera. This methodology is applicable to large testing facilities using multiple instruments.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。