Enumeration of CD34-positive Stem Cells Using the ADAMII Image-based Fluorescence Cell Counter

利用ADAMII基于图像的荧光细胞计数器对CD34阳性干细胞进行计数

阅读:1

Abstract

BACKGROUND: It is very important to accurately enumerate CD34-positive (CD34+) cells for successful hematopoietic stem cell transplantation (HSCT). We evaluated the ability of the newly developed image based-immunofluorescence cell counter ADAMII (NanoEntek, Seoul, Korea) to enumerate CD34+ cells, which was improved through simultaneous CD45 analysis. METHODS: We enumerated CD34+ cells with ADAMII using 19 peripheral blood (PB) and 91 leukapheresis samples from HSCT donors. Analytical performance, including precision and linearity, was analyzed, and sample stability during storage was evaluated. Viable CD34+ cell count (vCD34) and viable CD45+ cell count (vCD45) and the percentage of viable CD34+ cells among viable CD45+ cells (CD34/CD45) as measured by ADAMII were compared with the corresponding values from two flow cytometry assays, using regression analysis. RESULTS: ADAMII demonstrated acceptable precision, as CV values of vCD34 from six samples with different counts were all <10% (range: 3.49-9.51%). CV values of the vCD45 and CD34/45 ranged from 4.03% to 9.67% and from 2.48% to 10.07%, respectively. The linearity of vCD34 showed an excellent R² value (0.99) when analyzed using the intended count and flow cytometry data. The ADAMII and two flow cytometry-based assays generated very similar data for the PB and leukapheresis samples. CONCLUSIONS: ADAMII demonstrated excellent performance for use as a routine clinical assay in terms of CD34+ cell enumeration from PB and leukapheresis samples. Moreover, it could be used as a point-of-care-test for determining mobilization time and predicting an adequate apheresis stem cell product.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。