Performance of LIFECODES HLA-DQB1 typing kit using Luminex platform in Koreans

LIFECODES HLA-DQB1 分型试剂盒在韩国人群中使用 Luminex 平台的性能

阅读:1

Abstract

Intermediate-resolution HLA-DQ typing has gained importance in organ transplantation recently. We evaluated the performance of the LIFECODES HLA-DQB1 typing kit (Immucor, USA) using sequence-specific oligonucleotide (SSO) probe and Luminex platform (Luminex Corp., USA) on 100 samples tested by sequence-based typing (SBT) using the AlleleSEQR HLA-DQB1 kit (Abbott Molecular, USA) in Korean individuals. No sample showed ambiguity in the assignment of 4-digit HLA-DQB1 allele with the LIFECODES HLA-DQB1 SSO typing kit, and the results were fully concordant with those of high-resolution typing of AlleleSEQR HLA-DQB1 SBT up to 4-digit level. Three samples required adjustment of false reactions (3/100, 3.0%): two samples with DQB1(*)03:03/(*)06:01 showed false-positive result in probe 253, and 1 sample with DQB1(*)04:02/(*)05:02 showed false-negative result in probe 217. We tested an additional sample with DQB1(*)03:03/(*)06:01, which showed same false-positivity in probe 253 and 2 samples with DQB1(*)04:02/(*)05:02, which showed no false reaction. The false reactions did not result in ambiguity or change in the HLA allele assignment. We could assign HLA-DQB1 alleles to 4 digit-level without ambiguity, with 100% concordance with the SBT results. Thus, LIFECODES HLA-DQB1 SSO typing kit showed good performance for intermediate-resolution HLA-DQB1 typing in clinical laboratory for organ transplantation in Koreans.