Quartz crystal microbalances (QCM) are suitable for real-time dosimetry in nanotoxicological studies using VITROCELL®Cloud cell exposure systems

石英晶体微量天平 (QCM) 适用于使用 VITROCELL®Cloud 细胞暴露系统进行纳米毒理学研究中的实时剂量测定

阅读:6
作者:Yaobo Ding, Patrick Weindl, Anke-Gabriele Lenz, Paula Mayer, Tobias Krebs, Otmar Schmid

Background

Accurate knowledge of cell-/tissue-delivered dose plays a pivotal role in inhalation toxicology studies, since it is the key parameter for hazard assessment and translation of in vitro to in vivo dose-response. Traditionally, (nano-)particle toxicological studies with in vivo and in vitro models of the lung rely on in silio computational or off-line analytical

Conclusions

The QCMs of the Cloud 6 and Cloud 12 systems show similar performance and are highly sensitive, accurate devices for (quasi-) real-time dosimetry of the cell-delivered particle dose in ALI cell exposure experiments, if operated according to manufacturer specifications. Comparison with in vitro onset doses from this and previously published ALI studies revealed that the detection limit of 170 ng/cm2 is sufficient for determination of toxicological onset doses for all particle types with low (e.g. polystyrene) or high mass-specific toxicity (e.g. ZnO and Ag) investigated here. Hence, in principle QCMs are suitable for in vitro nanotoxciological studies, but this should be investigated for each QCM and ALI exposure system under the specific exposure conditions as described in the present study.

Results

Quantitative fluorescence spectroscopy of fluorescein-spiked saline aerosol was used to determine detection limit, precision and accuracy of the QCMs implemented in a VITROCELL®Cloud 6 and Cloud 12 system for dose-controlled ALI aerosol-cell exposure experiments. Both QCMs performed linearly over the entire investigated dose range (200 to 12,000 ng/cm2) with an accuracy of 3.4% (Cloud 6) and 3.8% (Cloud 12). Their precision (repeatability) decreased from 2.5% for large doses (> 9500 ng/cm2) to values of 10% and even 25% for doses of 1000 ng/cm2 and 200 ng/cm2, respectively. Their lower detection limit was 170 ng/cm2 and 169 ng/cm2 for the Cloud 6 and Cloud 12, respectively. Dose-response measurements with (NM110) ZnO nanoparticles revealed an onset dose of 3.3 μg/cm2 (or 0.39 cm2/cm2) for both cell viability (WST-1) and cytotoxicity (LDH) of A549 lung epithelial cells. Conclusions: The QCMs of the Cloud 6 and Cloud 12 systems show similar performance and are highly sensitive, accurate devices for (quasi-) real-time dosimetry of the cell-delivered particle dose in ALI cell exposure experiments, if operated according to manufacturer specifications. Comparison with in vitro onset doses from this and previously published ALI studies revealed that the detection limit of 170 ng/cm2 is sufficient for determination of toxicological onset doses for all particle types with low (e.g. polystyrene) or high mass-specific toxicity (e.g. ZnO and Ag) investigated here. Hence, in principle QCMs are suitable for in vitro nanotoxciological studies, but this should be investigated for each QCM and ALI exposure system under the specific exposure conditions as described in the present study.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。