NONINVASIVE TEAR FILM BREAK-UP TIME ASSESSMENT USING HANDHELD LIPID LAYER EXAMINATION INSTRUMENT

使用手持式脂质层检查仪进行无创泪膜破裂时间评估

阅读:1

Abstract

The aim was to determine feasibility and reliability of noninvasive tear break-up time (NIBUT) assessment using handheld lipid layer examination instrument, and to compare it with standard tear break-up time (TBUT) test. Fifty patients were enrolled, 31 with and 19 without dry eye symptoms. Schein questionnaire was used to assess dry eye symptoms. During examination, three NIBUT measurements were performed on each eye using handheld instrument, followed by three TBUT measurements. Receiver operating characteristic curves, sensitivity, specificity and logistic regression analysis were generated. Median NIBUT values were significantly shorter in dry eye symptom group than in control group in all three measurements (9, 8 and 8 s vs. 21, 22 and 21 s; p<0.001). TBUT values showed no significant difference between the groups in the first measurement (p=0.053), but the values were significantly shorter in dry eye symptom group in second and third measurements (p=0.020). The cutoff value to distinguish patients with symptoms of dry eye from control group was 12 seconds for NIBUT and 8 seconds for TBUT, with NIBUT having significantly higher sensitivity, specificity, area under the receiver operating characteristic curve and positive predictive value. NIBUT, measured by handheld lipid layer examination instrument, was superior to TBUT in detecting dry eye.