Abstract
BACKGROUND: We report the first single-centre experience in Africa with the Sensei X robotic navigation system in an unselected subset of patients with atrial fibrillation (AF). METHODS: Data were recorded prospectively of all consecutive patients who underwent robotically assisted catheter ablation therapy using the Sensei X robotic navigation system at the Christiaan Barnard Memorial Hospital, Cape Town, South Africa, from July 2009 to July 2010. Outcomes were defined at one and nine months. RESULTS: A total of 95 patients were included: 63% had only AF and 37% had AF plus atrial flutter. AF was of the persistent type in 81% of patients. The mean procedure, fluoroscopy and ablation times were 220.6 ± 89.6 min, 31.0 ± 20.4 min, and 61.3 ± 28.1 min, respectively. Both fluoroscopy and procedure times were significantly longer for the first 19 patients compared with the remaining 76 patients (43.5 ± 22.7 vs 27.8 ± 18.5 min and 274.7 ± 90.2 vs 207.1 ± 84.7 min, respectively, p = 0.002). The procedural endpoint of the study was successfully achieved in all patients. After one attempt, 27% were discharged from hospital off anti-arrhythmic drugs (AADs). At a median of nine months' follow up, 74% were AF-free off AADs, and 11% were AF-free on AADs, yielding a total freedom from AF of 84% without any redo procedures. Freedom from relapse after 1.12 procedures was 88%. CONCLUSION: The Sensei X robotic navigation system offers a safe and effective approach for the treatment of AF. There was a learning curve with regard to fluoroscopy and procedure time, after which point reduction in radiation exposure and operator strain, as well as improvement in procedure throughputs were even more pronounced.