Feasibility Testing of the RT Cardiac Systems Percutaneous Mechanical Circulatory Support Device

RT心脏系统经皮机械循环支持装置的可行性测试

阅读:1

Abstract

RT Cardiac Systems (RTCS, Raleigh, NC) is developing an intravascular percutaneous mechanical circulatory support (pMCS) device drive system for use during high-risk percutaneous coronary intervention and emergent cardiogenic shock. The proprietary pMCS device (US patent 10,780,206) consists of a miniaturized axial flow pump with an integrated motor connected via a short flexible drive system. This novel flexible drive system creates a flexible pump that is advantageous for percutaneous placement and conforming to anatomy. This design also has the benefit of not requiring a continuous external lubrication source. In this article, we present engineering development and feasibility testing of the prototype pMCS system. Computational fluid dynamics (CFD) modeling was performed to evaluate candidate blade set designs (impeller leading and trailing edges, diffuser) and predict hydrodynamic performance and hemolysis risk. Bench testing of candidate lip seal designs (radial interference, durometer, and seal angle) was evaluated for leak rate. Two 16Fr prototype devices were then fabricated and tested in a static mock flow loop. Experimental testing demonstrated 3 L/min flow against 110 mmHg and 4 L/min flow against 80 mmHg, which matched the CFD-predicted hydrodynamic performance. These results demonstrate feasibility of the engineering design and performance of the prototype devices.