Novel Left Atrial Cannulation Technique for Attachment of a Pumpless Artificial Lung

一种用于连接无泵人工肺的新型左心房插管技术

阅读:1

Abstract

A pumpless artificial lung has the potential to provide a bridge to recovery or transplantation in children with respiratory failure. Pulmonary artery inflow and left atrial outflow are necessary for low-gradient, pumpless systems; however, long-term cannulation of the fragile left atrium remains problematic. In this technique, the left atrium and pulmonary artery were exposed through a left anterior thoracotomy. Inflow to the artificial lung was created using an end-to-side anastomosis with the pulmonary artery. Device outflow was established through the left atrium. A single-stage venous cannula was passed through a free PTFE graft. Using polypropylene with pledgets, two concentric purse-string sutures were placed in the dome of the left atrium. The venous cannula was inserted. The graft was slid down the cannula and circumferentially secured to the adjacent left atrial tissue and pledgets. The other end of the graft was secured to the cannula with silk ties. The procedure was successful in 10 sheep. Initial device blood flow was 969 ± 222 ml/min, which remained stable for up to 7 days with no anastomotic complications. This is an effective method of achieving secure, long-term left atrial cannulation without cardiopulmonary bypass for use in a low-resistance, pumpless artificial lung. And, most importantly, improves the ease and safety of cannula replacement and final decannulation when AL support is no longer required.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。