Regional Ventilation-Perfusion Matching by Electrical Impedance Tomography After Single Lung Transplant

单肺移植后通过电阻抗断层扫描进行区域通气/灌注匹配

阅读:1

Abstract

Single lung transplantation (LUTX) can be the last therapeutic option for a growing cohort of patients suffering from end-stage respiratory failure. Postoperative ventilatory management of single LUTX recipients is challenged by the coexistence of the diseased native lung and a healthy-but fragile-graft. In this case report, in a single LUTX recipient with idiopathic pulmonary fibrosis, regional ventilation ( ), perfusion ( ), and / matching and subsequent measurement of shunt fraction ( Qs / Qt ) and alveolar dead space ( Vd / Vt ) were obtained by integrating electrical impedance tomography (EIT) with volumetric capnography and pulmonary thermodilution technique. Although the preoperative pulmonary scintigraphy showed predominant right lung perfusion (79.8% vs. 20.2%), the EIT documented the postoperative re-establishment of between the lungs (demonstrating the adequate functioning of vascular anastomoses), the diversion of to the graft and similar global Qs / Qt (17%) and Vd / Vt (29%) between native and graft lung. Electrical impedance tomography mapping allowed regional Qs / Qt and Vd / Vt assessment: the native right lung had a completely deranged distribution of and ( Qs / Qt 25%, Vd / Vt 46%), whereas the graft showed normal coupling of and ( Qs / Qt 8%, Vd / Vt 12%). Electrical impedance tomography may allow noninvasive, repeatable, bedside assessments of the lung / coupling after single LUTX.