Effect of Hematocrit and Elevated Beat Rate on the 12cc Penn State Pediatric Ventricular Assist Device

血细胞比容和心率增快对12cc宾夕法尼亚州立大学儿科心室辅助装置的影响

阅读:1

Abstract

Congenital heart disease affects approximately 40,000 infants annually in the United States with 25% requiring invasive treatment. Due to limited number of donor hearts and treatment options available for children, pediatric ventricular assist devices (PVADs) are used as a bridge to transplant. The 12cc pneumatic Penn State PVAD is optimized to prevent platelet adhesion and thrombus formation at patient nominal conditions; however, children demonstrate variable blood hematocrit and elevated heart rates. Therefore, with pediatric patients exhibiting greater variability, particle image velocimetry is used to evaluate the PVAD with three non-Newtonian hematocrit blood analogs (20%, 40%, and 60%) and at two beat rates (75 and 120 bpm) to understand the device's performance. The flow fields demonstrate a strong inlet jet that transitions to a solid body rotation during diastole. During systole, the rotation dissipates and reorganizes into an outlet jet. This flow field is consistent across all hematocrits and beat rates but at a higher velocity magnitude during 120 bpm. There are also minor differences in flow field timing and surface washing due to hematocrit. Therefore, despite patient differences in hematocrit or required pumping output, thorough surface washing can be achieved in the PVAD by altering operating conditions, thus reducing platelet adhesion potential.