Toward a Servoregulation Controller to Automate CO2 Removal in Wearable Artificial Lungs

面向可穿戴人工肺二氧化碳自动去除的伺服调节控制器

阅读:2

Abstract

A laptop-driven, benchtop control system that automatically adjusts carbon dioxide (CO2) removal in artificial lungs (ALs) is described. The proportional-integral-derivative (PID) feedback controller modulates pump-driven air sweep gas flow through an AL to achieve a desired exhaust gas CO2 partial pressure (EGCO2). When EGCO2 increases, the servoregulator automatically and rapidly increases sweep flow to remove more CO2. If EGCO2 decreases, the sweep flow decreases to reduce CO2 removal. System operation was tested for 6 hours in vitro using bovine blood and in vivo in three proof-of-concept sheep experiments. In all studies, the controller automatically adjusted the sweep gas flow to rapidly (<1 minute) meet the specified EGCO2 level when challenged with changes in inlet blood or target EGCO2 levels. CO2 removal increased or decreased as a function of arterial pCO2 (PaCO2). Such a system may serve as a controller in wearable AL systems that allow for large changes in patient activity or disease status.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。