First Implantation of Silicon Nanopore Membrane Hemofilters

首次植入硅纳米孔膜血液滤器

阅读:1

Abstract

An implantable hemofilter for the treatment of kidney failure depends critically on the transport characteristics of the membrane and the biocompatibility of the membrane, cartridge, and blood conduits. A novel membrane with slit-shaped pores optimizes the trade-off between permeability and selectivity, enabling implanted therapy. Sustained (3-8) day function of an implanted parallel-plate hemofilter with minimal anticoagulation was achieved by considering biocompatibility at the subnanometer scale of chemical interactions and the millimeter scale of blood fluid dynamics. A total of 400 nm-thick polysilicon flat sheet membranes with 5-8 nm × 2 micron slit-shaped pores were surface-modified with polyethylene glycol. Hemofilter cartridge geometries were refined based on computational fluid dynamics models of blood flow. In an uncontrolled pilot study, silicon filters were implanted in six class A dogs. Cartridges were connected to the cardiovascular system by anastamoses to the aorta and inferior vena cava and filtrate was drained to collection pouches positioned in the peritoneum. Pain medicine and acetylsalicylic acid were administered twice daily until the hemofilters were harvested on postoperative days 3 (n = 2), 4 (n = 2), 5 (n = 1), and 8 (n = 1). No hemofilters were thrombosed. Animals treated for 5 and 8 days had microscopic fractures in the silicon nanopore membranes and 20-50 ml of transudative (albumin sieving coefficient θalb ~ 0.5 - 0.7) fluid in the collection pouches at the time of explant. Shorter experimental durations (3-4 days) resulted in filtration volumes similar to predictions based on mean arterial pressures and membrane hydraulic permeability and (θalb ~ 0.2 - 0.3), similar to preimplantation measurements. In conclusion, a detailed mechanistic and materials science attention to blood-material interactions allows implanted hemofilters to resist thrombosis. Additional testing is needed to determine optimal membrane characteristics and identify limiting factors in long-term implantation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。