Development of an inlet pressure sensor for control in a left ventricular assist device

开发用于左心室辅助装置控制的入口压力传感器

阅读:1

Abstract

A Tesla type continuous flow left ventricular assist device (VAD) has been designed by Penn State and Advanced Bionics, Inc. (ABI). When a continuous flow device is used, care must be taken to limit low pressures in the ventricle, which can produce an obstruction to the inlet cannula or trigger arrhythmias. Design of an inexpensive, semiconductor strain gauge inlet pressure sensor to detect suction has been completed. The research and design analysis included finite element modeling of the sensing region. Sensitivity, step-response, temperature dependence, and hysteresis tests have been performed on prototype units. All sensors were able to withstand the maximum expected strain of 82 microm/in at 500 mm Hg internal pressure. Average sensitivity was 0.52 +/- 0.24 microV/mm Hg with 0.5 V excitation (n = 5 units). Step-response time for a 0- to 90-mm Hg step change averaged 22 msec. Hysteresis was measured by applying and holding 75 mm Hg internal pressure for 4 hours, followed by a zero pressure measurement, and ranged from -15 to 4.1 mm Hg (n = 3 units). Offset drift varied between 180 and -140 mm Hg over a 4-week period (n = 2 units). Span temperature sensitivity ranged from 18 to -21 muV/ degrees C (n = 5 units). Gain temperature sensitivity ranged from -7.4 to 4.9 muV/ degrees C (n = 5 units). With the inherent drift, it is currently not possible to use the transducer to measure actual pressures, but it can easily be used to measure pressure changes throughout the cardiac cycle. This signal can then be used in the control system to avoid ventricular suction events.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。