High-Sensitivity Determination of K, Ca, Na, and Mg in Salt Mines Samples by Atomic Emission Spectrometry with a Miniaturized Liquid Cathode Glow Discharge

利用小型化液阴极辉光放电原子发射光谱法高灵敏度测定盐矿样品中的钾、钙、钠和镁

阅读:1

Abstract

An atomic emission spectrometer (AES) based on a novel atmospheric pressure liquid cathode glow discharge (LCGD) as one of the most promising miniaturized excitation sources has been developed, in which the glow discharge is produced between a needle-like Pt anode and the electrolyte (as cathode) overflowing from a quartz capillary. Lower energy consumption (<50 W) and higher excitation efficiency can be realized by point discharge of the needle-like Pt. The miniaturized LCGD seems particularly well suited to rapid and high-sensitivity determination of K, Ca, Na, and Mg in salt mines samples. The optimized analytical conditions of LCGD-AES were pH = 1 with HNO(3) as electrolyte, 650 V discharge voltage, and 3 mL min(-1) solution flow rate. The limits of detections (LODs) of K, Ca, Na, and Mg were 0.390, 0.054, 0.048, and 0.032 mg L(-1), respectively. Measurement results of the LCGD-AES are in good agreement with the comparison value obtained by inductively coupled plasma (ICP) and ion chromatography (IC). All results suggested that the developed portable analytical instrument can be used for on-site and real-time monitoring of metal elements in field with further improvement.