Safety outcomes of the first Rigi10™ malleable penile prostheses implanted worldwide

全球首例Rigi10™可弯曲阴茎假体植入的安全结果

阅读:1

Abstract

Implantation of penile prosthesis is considered when conservative measures fail or are unacceptable to patients' wishing treatment for erectile dysfunction. In the United States (US), inflatable penile prostheses are more often used than malleable penile prostheses (MPP). Outside the US, the reverse is true because third-party reimbursement is not available, and MPP is considerably cheaper. Two American manufacturers make MPP; presently, a new manufacturer, Rigicon (Ronkonkoma NY), has recently begun to sell its MPP worldwide. Patient information forms submitted to the manufacturer between March 1, 2019, and December 8, 2022, were used to conduct an initial safety study for 605 first-time patients implanted with Rigicon10® by 46 physicians in 15 countries with a mean follow-up of 21.6 months. It has the same configuration of trimmable, paired silicone rods containing a twisted stainless-steel wire for bendability. However, it is available in six widths with hydrophilic coating compared to three widths offered by competitors. Revision or explantation was needed in 6 of 605 patients (0.99%) with half of those being removed for dissatisfaction (0.50%). Two (0.33%) suffered device infection and one (0.16%) required removal for erosion. Kaplan-Meier's statistical analysis showed three-year implant survival from revision = 99.2%. It demonstrated a comparable safety record with less than 1.00% of patients requiring reoperation.