Effect of serum content and diluent selection on assay sensitivity and signal intensity in multiplex bead-based immunoassays

血清含量和稀释剂选择对多重珠基免疫测定中测定灵敏度和信号强度的影响

阅读:7
作者:Christian Pfleger, Nanette Schloot, Frank ter Veld

Conclusions

In a 19-plex setup, this paper describes signal intensity, assay sensitivity and background signal levels in relation to the total volume-fraction of serum and protein concentration. For the determination of cytokines in serum samples with the multiplexed system Luminex the diluent RD6 seems more appropriate than the diluent DY997.

Methods

Two commercially available diluents, DY997 and RD6 from R&D Systems, were compared in a 19-plex immunoassay setup from Luminex.

Results

Using diluent DY997, multiplex signal intensity was reduced by 55% when spiked samples (chemokines and cytokines at 100 pg/mL) contained 50% v/v human serum, compared to samples containing 25% v/v. When using diluent RD6, signal intensity was reduced by 20% when samples contained 50% v/v human serum, compared to 25% v/v human serum. Diluent DY997 showed decreasing multiplex assay sensitivity with increasing protein concentrations, but not as low as in the presence of 50% v/v human serum. Conclusions: In a 19-plex setup, this paper describes signal intensity, assay sensitivity and background signal levels in relation to the total volume-fraction of serum and protein concentration. For the determination of cytokines in serum samples with the multiplexed system Luminex the diluent RD6 seems more appropriate than the diluent DY997.