Accuracy of postvoid residual volumes after vaginal delivery: a prospective equivalence study to compare an automatic scanning device with transurethral catheterization

阴道分娩后残余尿量测定的准确性:一项前瞻性等效性研究,比较自动扫描装置与经尿道导尿法

阅读:1

Abstract

INTRODUCTION AND HYPOTHESIS: Abnormal postvoid residual volumes (PVRV) after delivery are common in daily clinical practice. By using an automatic scanning device, unnecessary catheterizations can be prevented. The aim of this study was to determine the accuracy of PVRV after vaginal delivery measured by an automatic scanning device through a comparison with transurethral catheterization. MATERIALS AND METHODS: This prospective observational equivalence study was performed in patients who delivered vaginally between June 2012 and May 2017 in three teaching hospitals in The Netherlands. After the first spontaneous void after delivery, postvoid residual volume (PVRV) was measured with a portable automatic scanning device (BladderScan® BVI 9400). Directly afterward, it was measured by catheterization. Correlation between measurements was calculated using Spearman's correlation coefficient and agreement plot. The primary outcome was to validate the correlation between the BladderScan® compared with the gold standard of transurethral catheterization. RESULTS: Data of 407 patients was used for final analysis. Median PVRV as measured by BladderScan® was 380 ml (± 261-0-999 ml) and by catheterization was 375 ml (± 315-1800 ml). Mean difference between measurements was -12.9 ml (± 178 ml). There was a very good correlation between methods (Spearman's rho = 0.82, p < 0.001). Using a cut-ff value of >500 ml, specificity and sensitivity were 85.4 and 85.6%, respectively. CONCLUSIONS: The BladderScan® (BVI 9400) measures PVRV precisely and reliably after vaginal delivery and should be preferred over catheterization.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。