Quality Assurance Procedures based on Dosimetric, Gamma Analysis as a Fast Reliable Tool for Commissioning Brachytherapy Treatment Planning Systems

基于剂量学和伽马分析的质量保证程序,作为调试近距离放射治疗计划系统的快速可靠工具

阅读:1

Abstract

BACKGROUND: Fast and easily repeatable methods for commissioning procedures for brachytherapy (BT) treatment planning systems (TPS) are needed. Radiochromic film dosimetry with gamma analysis is widely used in external beam quality assurance (QA) procedures and planar film dosimetry is also increasingly used for verification of the dose distribution in BT applications. Using the gamma analysis method for comparing calculated and measured dose data could be used for commissioning procedures of the newly developed TG-186 and MBDCA calculation algorithms. The aim of this study was dosimetric verification of the calculation algorithm used in TPS when the CT/MRI ring applicator is used. MATERIALS AND METHODS: Ring applicators with 26 and 30 mm diameters and a 60 mm intra-uterine tube with 60° angle were used for verification. Gafchromic® EBT films were used as dosimetric media. Dose grids, corresponding to each plane (dosimetric film location), were exported from the TPS as a raw data. Gafchromic® films were digitized after irradiation. gamma analysis of the data were performed using the OMNI Pro I'mRT® system, as recommended by the AAPM TG-119 rapport criterion for gamma analysis of 3%, 3 mm and a level of 95%. RESULTS: For the 26 mm and 30 mm rings, the average gamma ranged, respectively, from 0.1 to 0.44 and from 0.1 to 0.27. In both cases, 99% of the measured points corresponded with the calculated data. CONCLUSIONS: This analysis showed excellent agreement between the dose distribution calculated with the TPS and the doses measured by Gafchromic films. This finding confirms the viability of using film dosimetry in BT.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。