Target and peripheral dose from radiation sector motions accompanying couch repositioning of patient coordinates with the Gamma Knife(®) Perfexion(™)

使用 Gamma Knife® Perfexion™ 治疗时,随着治疗床重新定位患者坐标,放射扇区运动引起的靶区和周边剂量

阅读:1

Abstract

BACKGROUND: The GammaPlan(™) treatment planning system (TPS) does not fully account for shutter dose when multiple shots are required to deliver a patient's treatment. The unaccounted exposures to the target site and its periphery are measured in this study. The collected data are compared to a similar effect from the Gamma Knife(®) model 4C. MATERIALS AND METHODS.: A stereotactic head frame was attached to a Leksell(®) 16 cm diameter spherical phantom; using a fiducial-box, CT images of the phantom were acquired and registered in the TPS. Measurements give the relationship of measured dose to the number of repositions with the patient positioning system (PPS) and to the collimator size. An absorbed dose of 10 Gy to the 50% isodose line was prescribed to the target site and all measurements were acquired with an ionization chamber. RESULTS: Measured dose increases with frequency of repositioning and with collimator size. As the radiation sectors transition between the beam on and beam off states, the target receives more shutter dose than the periphery. Shutter doses of 3.53±0.04 and 1.59±0.04 cGy/reposition to the target site are observed for the 16 and 8 mm collimators, respectively. The target periphery receives additional dose that varies depending on its position relative to the target. CONCLUSIONS: The radiation sector motions for the Gamma Knife(®) Perfexion(™) result in an additional dose due to the shutter effect. The magnitude of this exposure is comparable to that measured for the model 4C.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。