A novel polymer-free everolimus-eluting stent with a nitrogen-doped titanium dioxide film inhibits restenosis and thrombosis in a swine coronary model

一种新型的无聚合物依维莫司洗脱支架,其表面覆有氮掺杂二氧化钛薄膜,可在猪冠状动脉模型中抑制再狭窄和血栓形成。

阅读:2

Abstract

BACKGROUND: Short-term outcomes regarding the safety and efficacy of a polymer-free everolimus-eluting stent (EES) with a nitrogen-doped titanium dioxide (N-TiO₂) film in a swine coronary model have been reported. However, the long-term results of the use of this type of stent have not yet been evaluated or compared to those of other polymer-free coronary stents. Therefore, this study aimed to determine the mid- to long-term safety and efficacy of a polymer-free EES with an N-TiO₂ film in a swine coronary model. METHODS: Polymer-free EES with N-TiO₂ films (n = 30) and polymer-free sirolimus-eluting stents (SES; n = 30) were implanted in 30 pigs. Quantitative coronary analysis and optical coherence tomography were conducted immediately and at 1 (quantitative coronary analysis only), 3, and 6 months after stenting. Histopathologic examinations were performed at 1, 3, and 6 months after stenting. RESULTS: The polymer-free EES group had a lower percentage of neointimal growth than the polymer-free SES group at 3 months (22.5% ± 11.4% vs. 32.1% ± 12.3%; p < 0.001). The polymer-free EES group had a lower fibrin score than the polymer-free SES group at 1 month (1.9 ± 0.45 vs. 2.5 ± 0.54; p = 0.001). The re-endothelialization rates were similar between groups. The polymer-free EES group had a lower percentage of the area of stenosis than the polymer-free SES group throughout the follow-up period. CONCLUSIONS: The novel polymer-free EES with an N-TiO₂ film has superior safety and efficacy than the polymer-free SES at the 6-month follow-up in a swine model.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。