Implantation of the Micra transcatheter pacing system: Single Polish center experience with the real costs of hospitalization analysis

Micra经导管起搏系统植入术:波兰单一中心经验及住院实际成本分析

阅读:1

Abstract

BACKGROUND: The Micra transcatheter pacing system (TPS) is a miniaturized, single-chamber pacemaker system. Study reported herein is an initial experience with implantation of the Micra TPS. METHODS: The leadless pacemaker was implanted in 10 patients with standard indications for a permanent pacemaker implantation. All hospitalization costs were calculated for all patients. RESULTS: The mean age of the patients was 75 ± 7.1 years, 6 were men and 4 were women. Four patients had permanent atrial fibrillation as the basal rhythm and 6 patients had sinus rhythm. All patients had at least one relative contraindication that precluded the use of a traditional pacing system. Mean intraoperative ventricular sensing amplitude was 10.6 ± 5.4 mV, impedance 843 ± 185 ohms, and pacing threshold at 0.24 ms was 0.56 ± 0.23 V. At discharge, those values were 13.9 ± 5.6 mV, 667 ± 119 ohms and 0.47 ± 0.17, respectively. The mean duration of implantation procedure was 82 min, while mean fluoroscopy time was 3.5 min. Two patients developed hematoma at the groin puncture site post-implantation. In 1 case there was a need for erythrocyte mass transfusion and surgical intervention. Mean total time of hospitalization was 26 days and time from procedure to discharge 12 days. Average cost of hospitalization per 1 patient was 11,260.15 EUR minimal cost was 9,052.68 EUR, while maximal cost was 16,533.18 EUR. CONCLUSIONS: Implantation of leadless pacemakers is feasible, safe and provides advantages over the conventional system. Hospitalization costs vary for individual patients in wide range.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。