Two-year longitudinal evaluation of a second-generation thin-strut sirolimus-eluting bioresorbable coronary scaffold with hybrid cell design in porcine coronary arteries

对采用混合细胞设计的第二代细支架西罗莫司洗脱生物可吸收冠状动脉支架在猪冠状动脉中的两年纵向评估

阅读:1

Abstract

BACKGROUND: The first commercially available bioresorbable scaffold (BRS) had a strut thickness of 156 microns. As such, it had the potential for delivery challenges and higher thrombogenicity. The aim herein, is to evaluate biomechanical performance, pharmacokinetics and vascular healing of a novel thin strut (100 μm) sirolimus eluting BRS (MeRes-100, Meril Life Sciences, Gujarat, India) against the once clinically used BRS (Absorb BVS, Abbott, Santa Clara, CA) in porcine coronary arteries. METHODS: Following device implantation, angiographic and optical coherence tomography (OCT) evaluation were performed at 45, 90, 180 days, 1 year and 2 years. Histological evaluation was per-formed at 30, 90 and 180 days. RESULTS: At 2 years, both lumen (MeRes-100 7.07 ± 1.82 mm² vs. Absorb BVS 7.57 ± 1.39 mm2, p = NS) and scaffold areas (MeRes-100 9.73 ± 1.80 mm² vs. Absorb BVS 9.67 ± 1.25 mm², p = NS) were comparable between tested and control scaffolds. Also, the late lumen area gain at 2 years was similar in both groups tested (MeRes-100 1.03 ± 1.98 mm² vs. Absorb BVS 0.85 ± 1.56 mm², p = NS). Histologic examination up to 6 months showed comparable healing and inflammation profiles for both devices. CONCLUSIONS: The novel sirolimus-eluting BRS with thinner struts and hybrid cell design showed similar biomechanical durability and equivalent inhibition of neointimal proliferation when compared to the first-ever Absorb BVS up to 2 years in normal porcine coronary arteries.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。