Accidental Magnetic Resonance Imaging Activation of Carbon Dioxide Tissue Expanders

意外磁共振成像激活二氧化碳组织扩张器

阅读:1

Abstract

Implant-based reconstruction is the most common form of breast reconstruction following mastectomy. It is most often performed in 2 stages using saline-based tissue expanders, which are then exchanged for permanent implants. Serial expansions are performed by accessing a port in the office, an inconvenient and sometimes painful process. A carbon dioxide tissue expander is a device that provides a needle-free, patient-controlled expansion utilizing a remote-controlled CO(2) canister. While a patient-controlled expansion offers convenience, given that the CO(2) reservoir holds approximately 1500 mL of gas, the potential for malfunction resulting in an uncontrolled expansion in unique to this device. The authors present a case report of a patient with bilateral pre-pectoral tissue expanders who underwent magnetic resonance imaging, resulting in uncontrolled expansion.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。