The Use of Catheter Mount Will Result in More Reliable Carbon Dioxide Monitoring under Fluid Exposing Conditions

使用导管固定装置将使在液体暴露条件下的二氧化碳监测更加可靠。

阅读:2

Abstract

INTRODUCTION: Capnometer can be readily malfunctioned by fluid exposure during treatment of critically ill patients. This study aimed to determine whether placing capnometer distant from the endotracheal tube by connecting direct connect catheter mount (DCCM) is effective in yielding reliable end-tidal carbon dioxide (ETCO(2)) by reducing capnometer malfunctioning caused by water exposure. METHODS: In 25 healthy adults, a prospective, open label, crossover study was conducted to examine the effect of DCCM in mainstream and microstream capnometers under water exposing conditions. The primary endpoint was the comparison of ETCO(2) between proximal DCCM (pDCCM) and distal DCCM (dDCCM). RESULTS: For mainstream capnometers, mean ETCO(2) was significantly (p < 0.001) higher in dDCCM compared to pDCCM under water exposing conditions (29.5 vs. 19.0 with 5 ml; 33.8 vs. 21.2 with 10 ml; mmHg). Likewise, for microstream capnometers, ETCO(2) was greatly higher (p < 0.001) in dDCCM compared to pDCCM (30.5 vs. 13.9 with 5 ml; 29.9 vs.11.4 with 10 mL; mmHg). ETCO(2) measured by dDCCM was reliable in microstream settings, whereas it was unreliable in mainstream (correlation coefficient 0.88 vs. 0.27). CONCLUSIONS: Application of DCCM onto the capnometer setting seems to be effective in reducing capnometer malfunctioning under fluid exposing conditions, which is obvious in microstream capnometer by producing more reliable ETCO(2).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。