RPLC Determination of Tinidazole and Diloxanide Furoate in Tablets

反相高效液相色谱法测定片剂中替硝唑和糠酸二氯尼特的含量

阅读:1

Abstract

A simple reverse phase liquid chromatographic method has been developed and subsequently validated for simultaneous determination of tinidazole and diloxanide furoate. The separation was carried out using a mobile phase consisting of acetonitrile, methanol and 0.2 M potassium dihydrogen phosphate (pH 5) in the ratio 2:3:2.The column used was SS Wakosil-II C-18 with a flow rate of 1 ml/min and UV detection at 282 nm. The described method was linear over a concentration range of 10-70 μg/ml and 10-90 μg/ml for the assay of diloxanide furoate and tinidazole, respectively. The mean recovery was found to be 100-101% for tinidazole and 97-103% for diloxanide furoate when determined at three different levels.