Abstract
A simple reverse phase liquid chromatographic method has been developed and subsequently validated for simultaneous determination of salbutamol sulphate and bromhexine hydrochloride. The separation was carried out using a mobile phase consisting of acetonitrile, methanol and phosphate buffer, pH 4 in the ratio 60:20:20 v/v. The column used was SS Wakosil-II C-18 with a flow rate of 1 ml/min and UV detection at 224 nm. The described method was linear over a concentration range of 10-110 mug/ml and 20-140 mug/ml for the assay of salbutamol sulphate and bromhexine hydrochloride, respectively. The mean recovery was found to be 95-105% for salbutamol sulphate and 96.2-102.1% for bromhexine hydrochloride when determined at five different levels.