Balloon-Inflated Catheters for Enteral Feeding: a Word of Caution

用于肠内营养的球囊充气导管:注意事项

阅读:1

Abstract

Catheters with inflatable balloons such as a Foley catheter may be used for feeding gastrostomy/jejunostomy. The incorrect or improper use of these catheters can have serious consequences. We report 13 cases of feeding jejunostomy with balloon-inflated catheter's malfunction, some referred to our centre and others operated here over a period of 8 years. The most dramatic consequence of such improper use led to rupture of the small intestine due to inadvertent over-inflation (over 100 ml) of the balloon of the catheter during a contrast study. The patient required a laparotomy with resection and anastomosis of the bowel. Three other patients had similar over-inflation of the balloon leading to severe pain and discomfort. In all three patients, timely deflation of the balloon was sufficient to relieve the symptoms. One patient had intussusception with the inflated balloon acting as a lead point. The patient underwent resection of the small bowel with end jejunostomy and distal mucous fistula. All other patients presented with abdominal pain and distension and intestinal obstruction and were managed non-operatively with deflation of balloon either by aspiration, cutting the balloon port or ultrasound-guided puncture of balloon. Healthcare personnel dealing with patients with indwelling catheters must be educated to suspect, detect and manage such problems. The best measure for such unusual complications of otherwise safe devices would be prevention by training and generation of awareness.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。