Laparoscopic stereoscopic augmented reality: toward a clinically viable electromagnetic tracking solution

腹腔镜立体增强现实:迈向临床可行的电磁跟踪解决方案

阅读:1

Abstract

The purpose of this work was to develop a clinically viable laparoscopic augmented reality (AR) system employing stereoscopic (3-D) vision, laparoscopic ultrasound (LUS), and electromagnetic (EM) tracking to achieve image registration. We investigated clinically feasible solutions to mount the EM sensors on the 3-D laparoscope and the LUS probe. This led to a solution of integrating an externally attached EM sensor near the imaging tip of the LUS probe, only slightly increasing the overall diameter of the probe. Likewise, a solution for mounting an EM sensor on the handle of the 3-D laparoscope was proposed. The spatial image-to-video registration accuracy of the AR system was measured to be [Formula: see text] and [Formula: see text] for the left- and right-eye channels, respectively. The AR system contributed 58-ms latency to stereoscopic visualization. We further performed an animal experiment to demonstrate the use of the system as a visualization approach for laparoscopic procedures. In conclusion, we have developed an integrated, compact, and EM tracking-based stereoscopic AR visualization system, which has the potential for clinical use. The system has been demonstrated to achieve clinically acceptable accuracy and latency. This work is a critical step toward clinical translation of AR visualization for laparoscopic procedures.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。