Fabrication and assessment of a bio-inspired synthetic tracheal tissue model for tracheal tube cuff leakage testing

仿生合成气管组织模型的制备与评估及其在气管插管套囊泄漏测试中的应用

阅读:1

Abstract

INTRODUCTION: Leakage of orogastric secretions past the cuff of a tracheal tube is a contributory factor in ventilator-associated pneumonia. Current bench test methods specified in the International Standard for Anaesthetic and Respiratory Equipment (EN ISO 5361:2023) to test cuff leakage involve using a glass or plastic rigid cylinder model of the trachea. There is a need for more realistic models to inform cuff leakage. METHODS: We used human computerised tomography data and additive manufacturing (3D printing), combined with casting techniques to fabricate a bio-inspired synthetic tracheal model with analogous tissue characteristics. We conducted cuff leakage tests according to EN ISO 5361:2023 and compared results for high-volume low-pressure polyvinyl chloride and polyurethane cuffs between the rigid cylinder trachea with our bio-inspired model. RESULTS: The tracheal model demonstrated close agreement with published tracheal tissue hardness for cartilaginous and membranous soft tissues. For high-volume low-pressure polyvinyl chloride cuffs the leakage rate was >50% lower in the bio-inspired tracheal model compared with the rigid cylinder model (151 [8] vs 261 [11] ml h(-1)). For high-volume low-pressure polyurethane cuffs, much lower leakage rates were observed than polyvinyl chloride cuffs in both models with leakage rates higher for the bio-inspired trachea model (0.1 [0.2] vs 0 [0] ml h(-1)). CONCLUSION: A reproducible tracheal model that incorporates the mechanical properties of the human trachea can be manufactured from segmented CT images and additive manufactured moulds, providing a useful tool to inform future cuff development, leakage testing for industrial applications, and clinical decision-making. There are differences between cuff leakage rates between the bio-inspired model and the rigid cylinder recommended in EN ISO 5361:2023. The bio-inspired model could lead to more accurate and realistic cuff leakage rate testing which would support manufacturers in refining their designs. Clinicians would then be able to choose better tracheal tubes based on the outcomes of this testing.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。