Discordance between testosterone measurement methods in castrated prostate cancer patients

去势前列腺癌患者睾酮测量方法之间的差异

阅读:1

Abstract

Failure to suppress testosterone below 0.7 nM in castrated prostate cancer patients is associated with poor clinical outcomes. Testosterone levels in castrated patients are therefore routinely measured. Although mass spectrometry is the gold standard used to measure testosterone, most hospitals use an immunoassay method. In this study, we sought to evaluate the accuracy of an immunoassay method to measure castrate testosterone levels, with mass spectrometry as the reference standard. We retrospectively evaluated a cohort of 435 serum samples retrieved from castrated prostate cancer patients from April to September 2017. No follow-up of clinical outcomes was performed. Serum testosterone levels were measured in the same sample using liquid chromatography coupled with tandem mass spectrometry and electrochemiluminescent immunoassay methods. The mean testosterone levels were significantly higher with immunoassay than with mass spectrometry (0.672 ± 0.359 vs 0.461 ± 0.541 nM; P < 0.0001). Half of the samples with testosterone ≥0.7 nM assessed by immunoassay were measured <0.7 nM using mass spectrometry. However, we observed that only 2.95% of the samples with testosterone <0.7 nM measured by immunoassay were quantified ≥0.7 nM using mass spectrometry. The percentage of serum samples experiencing testosterone breakthrough at >0.7 nM was significantly higher with immunoassay (22.1%) than with mass spectrometry (13.1%; P < 0.0001). Quantitative measurement of serum testosterone levels >0.7 nM by immunoassay can result in an inaccurately identified castration status. Suboptimal testosterone levels in castrated patients should be confirmed by either mass spectrometry or an immunoassay method validated at low testosterone levels and interpreted with caution before any changes are made to treatment management.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。