Effects of Stimulus Duration on Event-Related Potentials Recorded From Cochlear-Implant Users

刺激持续时间对人工耳蜗植入者记录的事件相关电位的影响

阅读:1

Abstract

OBJECTIVES: Several studies have investigated the feasibility of using electrophysiology as an objective tool to efficiently map cochlear implants. A pervasive problem when measuring event-related potentials is the need to remove the direct-current (DC) artifact produced by the cochlear implant. Here, we describe how DC artifact removal can corrupt the response waveform and how the appropriate choice of stimulus duration may minimize this corruption. DESIGN: Event-related potentials were recorded to a synthesized vowel /a/ with a 170- or 400-ms duration. RESULTS: The P2 response, which occurs between 150 and 250 ms, was corrupted by the DC artifact removal algorithm for a 170-ms stimulus duration but was relatively uncorrupted for a 400-ms stimulus duration. CONCLUSIONS: To avoid response waveform corruption from DC artifact removal, one should choose a stimulus duration such that the offset of the stimulus does not temporally coincide with the specific peak of interest. While our data have been analyzed with only one specific algorithm, we argue that the length of the stimulus may be a critical factor for any DC artifact removal algorithm.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。