Setup Error and Effectiveness of Weekly Image-Guided Radiation Therapy of TomoDirect for Early Breast Cancer

TomoDirect每周图像引导放射治疗早期乳腺癌的设置误差和疗效

阅读:1

Abstract

PURPOSE: This study investigated setup error and effectiveness of weekly image-guided radiotherapy (IGRT) of TomoDirect for early breast cancer. MATERIALS AND METHODS: One hundred and fifty-one breasts of 147 consecutive patients who underwent breast conserving surgery followed by whole breast irradiation using TomoDirect in 2012 and 2013 were evaluated. All patients received weekly IGRT. The weekly setup errors from simulation to each treatment in reference to chest wall and surgical clips were measured. Random, systemic, and 3-dimensional setup errors were assessed. Extensive setup error was defined as 5 mm above the margin in any directions. RESULTS: All mean errors were within 3 mm of all directions. The mean angle of gantry shifts was 0.6°. The mean value of absolute 3-dimensional setup error was 4.67 mm. In multivariate analysis, breast size (odds ratio, 2.82; 95% confidence interval, 1.00 to 7.90) was a significant factor for extensive error. The largest significant deviation of setup error was observed in the first week of radiotherapy (p < 0.001) and the deviations gradually decreased with time. The deviation of setup error was 5.68 mm in the first week and within 5 mm after the second week. CONCLUSION: In this study, there was a significant association between breast size and significant setup error in breast cancer patients who received TomoDirect. The largest deviation occurred in the first week of treatment. Therefore, patients with large breasts should be closely observed on every fraction and fastidious attention is required in the first fraction of IGRT.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。