Acute intraocular toxicity caused by perfluorocarbon liquids: safety control systems of medical devices

全氟碳液体引起的急性眼内毒性:医疗器械安全控制系统

阅读:1

Abstract

BACKGROUND: Acute retinal toxicity has been demonstrated to be associated with the intraoperative use of perfluorocarbon liquids (PFCLs), especially perfluorooctane (PFO). Recently, several cases of PFO-associated blindness have been reported in Spain, Holland, France, Italy, the Middle East, and South America. METHODS: As a result, a new ISO guideline (ISO 16672:2020) was drafted, discussed, approved, and released in 2019. This recent ISO16672:2020 guideline recommends performing direct cytotoxicity tests as an option along with chemical analysis to measure PFCL quality (purity and safety). RESULTS: In this review paper, it has been emphasized why an appropriate biological test, specifically direct exposure of PFCL to live cells, for measuring cytotoxicity must be performed with each PFCL batch along with chemical analysis. CONCLUSIONS: The paper intends to compile all available information to discuss possible approaches for avoiding adverse clinical cases in future.