The analysis of tear meniscus parameters during daily soft contact lens wear using optical coherence tomography

利用光学相干断层扫描技术分析每日佩戴软性隐形眼镜期间的泪液弯月面参数

阅读:1

Abstract

PURPOSE: To evaluate tear meniscus parameters in soft contact lens wearers (SCL) using optical coherence tomography (OCT) and ImageJ software. METHODS: This prospective study included 50 soft contact lens wearers (group 1: 25 symptomatic SCL wearers (SCLW), group 2: 25 asymptomatic SCL wearers (ASCW)) and 25 healthy non-CL wearers (group 3 (NCLW)). SCLs were fitted on each eye of CL wearers, and the lower tear meniscus was imaged using OCT before CL insertion, immediately afterward, and reimaged 2, 5 and 10 h after insertion. Tear meniscus parameters, including tear meniscus height (TMH), depth (TMD), turbidity, and percentage area occupied by particles (PAOP) were measured in all groups. RESULTS: Turbidity and PAOP measurements at baseline in SCLW were significantly higher than in other groups (p < 0.05). There was no significant difference between TMH, TMD, turbidity, and PAOP parameters calculated at baseline visit and two hours after SCL insertion in all groups (p > 0.05 for 2 comparisons). The symptomatic SCL users had a significant decrease in TMH and TMD in the fifth hour. The turbidity and PAOP measurements of SCLW and ASCW at the fifth and tenth hours were significantly higher than those of NCLW (p < 0.05). CONCLUSION: TMD and height TMH decrease throughout the day in all participants; however, a significant decrease in these parameters was observed only in symptomatic SCL users at the fifth hour, at the earliest. As the duration of CL wear increases, turbidity and PAOP even in asymptomatic SCL wearers become significantly higher than those in healthy non-CL wearers. KEY MESSAGES: What Is Known • Contact lens wear is associated with an increased risk of dry eye. • Tear volume decreases gradually during contact lens wear. What Is New • Tear meniscus turbidity and particle area occupied by particles (PAOP) were higher in symptomatic contact lens wearers and they increase gradually during contact lens wear. • Tear meniscus turbidity and PAOP may be measures of how well the tear film and meniscus are functioning in contact lens wearers.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。