BioRad BioPlex® HIV Ag-Ab assay: Incidence of false positivity in a low-prevalence population and its effects on the current HIV testing algorithm

BioRad BioPlex® HIV 抗原抗体检测:低流行率人群中假阳性发生率及其对当前 HIV 检测流程的影响

阅读:1

Abstract

BACKGROUND: The BioPlex® HIV Ag-Ab assay, unlike other HIV 1/2 antigen/antibody immunoassays, is capable of differentiating positive HIV-1 antibodies (Groups M and O) from HIV-2 antibodies and/or HIV-1 p24 antigen in a single test. OBJECTIVE: The Alaska State Virology Laboratory (ASVL) adopted the BioPlex® HIV Ag-Ab assay early 2017 and can report on its performance in terms of false positivity in a low-prevalence population and its effects on the current HIV testing algorithm recommended by the Centers for Disease Control and Prevention (CDC). STUDY DESIGN: Specimens received between March 2017 and August 2018 were screened using the BioPlex® HIV Ag-Ab assay. Specimens screening positive for HIV antibodies or antigen were further confirmed using the Geenius™ HIV 1/2 Supplemental Assay and/or HIV RNA testing. RESULTS: Of the 12,338 sera screened by the BioPlex assay for HIV, 35 specimens were positive. Only 22 of the specimens were confirmed by supplemental testing and were considered to be truly positive (PPV, 62.9%). RNA was not detected in these cases suggesting initial false positivity on the BioPlex® HIV Ag-Ab assay. True positive results had index values (IDX) of >180 whereas false positive IDX's were between 1 and 4, with the exception of one specimen. CONCLUSIONS: We suggest that specimens demonstrating positivity with low IDX values <4 on the BioPlex® HIV Ag-Ab assay proceed directly to RNA testing, essentially bypassing supplemental antibody confirmation tests, to reduce turnaround time and cost of HIV confirmation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。