Quantification of complex precore mutations of hepatitis B virus by SimpleProbe real time PCR and dual melting analysis

利用SimpleProbe实时PCR和双熔解曲线分析法定量乙型肝炎病毒复杂前核心区突变

阅读:1

Abstract

BACKGROUND: Hepatitis B virus (HBV) precore G1896A mutation is associated with Hepatitis B e antigen (HBeAg) seroconversion. This mutation and the adjacent G1899A mutation also appear to associate with increased risk of hepatocellular carcinoma. Quantitative mutant dynamics may help determine the potential of these mutants as clinical biomarkers. However, a reliable method to quantify either mutant is not available, partly because the viral genome has polymorphisms in general and the precore mutations are complex. OBJECTIVES: (1) To develop a reliable and ultrasensitive assay for the quantification of HBV G1896A and/or G1899A mutants. (2) To obtain preliminary data on the quantities of the precore mutants in patients. STUDY DESIGN: A SimpleProbe real time PCR assay was developed to quantify the HBV precore mutants. Dual melting analysis and a primer-probe partial overlap approach were used to increase detection accuracy. A wild-type selective PCR blocker was also developed to increase mutant detection sensitivity. RESULTS: The assay correctly identified the precore sequence from all 62 patient samples analyzed. More than 97% of precore sequences in the GenBank can be recognized. Mutant detection sensitivity reached 0.001% using a wild type-selective PCR blocker. At least one precore mutant can be detected from all 20 HBeAg-positive individuals who were negative for precore mutations by DNA sequencing. CONCLUSIONS: The reliability of this ultrasensitive mutation quantification assay was demonstrated. The same approaches may be useful for the detection of other clinically significant mutations. Evolution of the precore mutants warrants further studies.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。