A flow-through cell counting assay for point-of-care enumeration of CD4 T-cells

用于即时计数 CD4 T 细胞的流通式细胞计数分析

阅读:4
作者:Simon Bystryak, Rajiv P Bandwar, Rasa Santockyte

Abstract

CD4 T-cell count is a priority for staging HIV disease and guiding clinical management as part of HIV care. Conventional CD4 T-cell enumeration methods based on flow cytometry are expensive, require well-trained personnel, and are challenging to use in rural, resource-scarce areas. A simple CD4 T-cell count test that can be used at point-of care, the Flow-Through cell Counting Assay (FTCA), is described in this article. The FTCA is based on the use of: 1) a special membrane that selectively retains white blood cells (WBCs); 2) a sample delivery system; and 3) optical signal detection. To show the feasibility of the FTCA, a proof-of-concept prototype of the FTCA cassette and digital camera or handheld reflectance meter were used for obtaining quantitative assay results within 30 min. The results show that the FTCA allows for quantitative enumeration of CD4 T-cells in the clinically relevant range of CD4 T-cell concentrations. The advantages of the FTCA technology, including simplicity, short analysis time, and portability, suggest that FTCA has great potential for use in clinical practice and wide applicability for other cell-based diagnostic tests.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。