Abstract
A new automated particle counting and sizing system was used in this investigation. The Electrozone/Celloscope® counter electronically detects the passage of red blood cells in a saline suspension. It was used to automatically plot the size distribution curves for graphically displaying the effects of sequentially increased hemolysis levels in blood samples. In this study the size distributions and counts of red blood cells in a saline solution of various concentrations (0.85%, 0.75%, 0.65%, 0.60%, 0.55%, 0.50%, 0.45%, 0.40%, 0.35%, 0.30%, 0.20%, 0.10%) were analyzed and compared to normal values. The accuracy and reproducibility of counts with the electronic counter are compared to hemocytometer counts. Some of the advantages and disadvantages of both systems are discussed. Normal size distribution curves first shift slightly to a larger particle size due to osmotic swelling (in 0.65%, 0.60%, 0.55% of saline solution). Upon further reduction of the saline concentration to .45%, .40%, .35%, hemolysis is observed as generation of the second peak showing clearly the smaller particle size.