A novel approach to mapping load transfer from the plantar surface of the foot to the walls of the total contact cast: a proof of concept study

一种绘制足底至全接触石膏壁载荷传递图的新方法:概念验证研究

阅读:1

Abstract

BACKGROUND: Total contact casting is regarded as the gold standard treatment for plantar foot ulcers. Load transfer from the plantar surface of the foot to the walls of the total contact cast has previously been assessed indirectly. The aim of this proof of concept study was to determine the feasibility of a new method to directly measure the load between the cast wall and the lower leg interface using capacitance sensors. METHODS: Plantar load was measured with pedar® sensor insoles and cast wall load with pliance® sensor strips as participants (n=2) walked along a 9 m walkway at 0.4±0.04 m/sec. The relative force (%) on the cast wall was calculated by dividing the mean cast wall force (N) per step by the mean plantar force (N) per step in the shoe-cast condition. RESULTS: The combined average measured load per step upon the walls of the TCC equated to 23-34% of the average plantar load on the opposite foot. The highest areas of load on the lower leg were located at the posterior margin of the lateral malleolus and at the anterior ankle/extensor retinaculum. CONCLUSIONS: These direct measurements of cast wall load are similar to previous indirect assessment of load transfer (30-36%) to the cast walls. This new methodology may provide a more comprehensive understanding of the mechanism of load transfer from the plantar surface of the foot to the cast walls of the total contact cast.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。