Brain perfusion CT compared with15O-H2O-PET in healthy subjects

健康受试者脑灌注CT与15O-H2O-PET的比较

阅读:1

Abstract

BACKGROUND: Regional cerebral blood flow [rCBF] measurements are valuable for identifying angiogenically active tumours, and perfusion computed tomography [CT] has been suggested for that purpose. This study aimed to validate rCBF measurements by perfusion CT with positron-emission tomography [PET] and15O-labelled water [15O-H2O] in healthy subjects. METHODS: RCBF was measured twice in 12 healthy subjects with15O-H2O PET and once with perfusion CT performed over the basal ganglia. Matching rCBF values in regions of interest were compared. RESULTS: Measured with perfusion CT, rCBF was significantly and systematically overestimated. White matter rCBF values were 17.4 ± 2.0 (mean ± SD) mL min-1 100 g-1 for PET and 21.8 ± 3.4 mL min-1 100 g-1 for perfusion CT. Grey matter rCBF values were 48.7 ± 5.0 mL min-1 100 g-1 for PET and 71.8 ± 8.0 mL min-1 100 g-1 for perfusion CT. The overestimation of grey matter rCBF could be reduced from 47% to 20% after normalization to white matter rCBF, but the difference was still significant. CONCLUSION: RCBF measured with perfusion CT does contain perfusion information, but neither quantitative nor relative values can substitute rCBF measured by15O-H2O PET yet. This, however, does not necessarily preclude a useful role in patient management.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。