Management of a remnant electrode in a patient with cardioverter-defibrillator infection after refusal of intravascular electrode removal

患者因植入式心脏复律除颤器感染而拒绝取出血管内电极,如何处理残留电极

阅读:1

Abstract

Treatments of choice for cardiac implantable electronic device (CIED) infections are the removal of the entire CIED system, control of infection, and new device implantation. Occasionally, a complete CIED removal can not be performed for several reasons, such as very old age, severe comobidity, limited life expectancy, or refusal by a patient. We encountered a male patient who developed traumatic CIED infection five years after cardioverter-defibrillator implantation. An intravenous electrode could not be removed by a simple transvenous extraction procedure, and he refused surgical removal of the remnant electrode. After control of local infection, the tips of the electrode were separated and buried between muscles, and the wound was closed with a local flap. CIED infection did not recur for 12 months even without relying on long-term antimicrobial treatment.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。