Abstract
INTRODUCTION: It is critical to measure the maximum voluntary bite force of patients receiving restorative dentistry. A new device known as "BYTE" has been developed indigenously to measure bite force in humans. The purpose of this study is to evaluate the BYTE device's consistency and accuracy in a lab setting. METHODOLOGY: Testing and calibration were done in the laboratory. The calibration machine with load cell pressed the biting part of the device with various forces from 3 N to 444 N in 3 N increments for two to three seconds each. The recorded force value in Newton by the device was noted down. RESULTS: At numerous standard loads, the minimum accuracy error is 0.333 N, while the maximum is 1.667 N. It marginally underestimates the load with an average accuracy error of 0.833 N. CONCLUSION: The calibration report showed that the BYTE device is precise and reliable and can be used to measure maximum bite force.