Assessment of Factors Affecting the Usefulness and Diagnostic Yield of Core Biopsy Needles with a Side Hole in Endoscopic Ultrasound-Guided Fine-Needle Aspiration

评估影响内镜超声引导下细针穿刺术中带侧孔穿刺针的有效性和诊断率的因素

阅读:1

Abstract

BACKGROUND/AIMS: A barbed puncture needle with a side hole was recently developed to improve sample quality and quantity in endoscopic ultrasound-guided fine-needle aspiration (EUS-FNA). In this study, we retrospectively assessed the usefulness of this puncture needle. METHODS: Factors affecting diagnostic yield, safety, and diagnostic accuracy were investigated in 76 patients who consecutively underwent EUS-FNA for neoplastic lesions at our hospital between January and December 2013. RESULTS: The procedure was successful in all cases; the rates of sample collection and determination of the correct diagnosis were 92.1% and 89.5%, respectively. The mean number of needle passes required for diagnosis was 1.1. Complications included mild intraluminal bleeding in two patients (2.6%). Multivariate analysis revealed that lesion size (≤20 mm) was significantly associated with a decreased chance of determining the correct diagnosis. CONCLUSIONS: Core biopsy needles with a side hole are safe and provide a satisfactory diagnostic yield. However, the side hole may potentially reduce the rate of making the correct diagnosis in small lesions.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。