Feasibility of newly developed endoscopic ultrasound with zone sonography technology for diagnosis of pancreatic diseases

新型内镜超声结合区域超声成像技术在胰腺疾病诊断中的可行性研究

阅读:1

Abstract

BACKGROUND/AIMS: To confirm the feasibility of using newly developed endoscopic ultrasound (EUS) with Zone sonography™ technology (ZST; Fujifilm Corp.). METHODS: Seventy-five patients with pancreatic disorders were enrolled: 45 with intraductal papillary mucinous neoplasm; 15 with ductal carcinoma; five with neuroendocrine tumors; three with serous cystic neoplasms; and seven with simple cysts. The endoscopes used were EG-530UR2 and EG-530UT2 (Fujifilm Corp.). Two items were evaluated: visualization depth among four frequencies and image quality after automatic adjustment of sound speed (AASS), assessed using a 5-scale Likert scale by two endosonographers blinded to disease status. Because sound speed could be manually controlled, besides AASS, image quality at sound speeds of 1,440 and 1,600 m/sec were also assessed. RESULTS: In all cases, sufficient images were obtained in the range of 3 cm from the EUS probe. Judgments of image quality before AASS were 3.49±0.50, 3.65±0.48, respectively. After AASS, A and B scored 4.36±0.48 and 4.40±0.49 (p<0.0001). There were significant differences in the data before and after AASS and plus 60 m/sec, but no significant difference between the datasets were seen after AASS and at sound speeds manually set for minus 100 m/sec. CONCLUSIONS: EUS with ZST was shown to be feasible in this preliminary experiment. Further evaluation of this novel technology is necessary and awaited.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。