The relationship between electrically evoked compound action potential and speech perception: a study in cochlear implant users with short electrode array

电诱发复合动作电位与言语感知的关系:一项针对短电极阵列人工耳蜗植入者的研究

阅读:1

Abstract

OBJECTIVES: To determine the extent to which electrically evoked compound action potential (ECAP) measurements were related with speech perception performance in implant users with a short electrode array and to investigate the relationship between ECAP measures and performance according to specific devices. DESIGN: Prospective study. SETTING: Tertiary referral center. PATIENTS: Seventeen Hybrid cochlear implant users were tested in this study. Subjects were divided into 2 groups: 8 using the Nucleus Hybrid M and 9 using the Nucleus Hybrid RE. In addition, 21 Nucleus Freedom long electrode implant (CI24RE) users also were tested to compare with the results of the old device (CI24M). MAIN OUTCOME MEASURES: ECAP growth functions were recorded using either an interphase gap (IPG) of 8 or 45 mus. We then calculated the slope of the growth function and changes in sensitivity with IPG. For each subject, these measures were compared with performance on tests of word recognition. RESULTS: The changes in sensitivity using 2 IPGs showed no correlation with the results of word recognition test in Hybrid cochlear implant users. In contrast, relatively strong correlations have been found between the slope of ECAP growth functions and performance on word recognition test. Additionally, when we separate the results of Hybrid M and RE, the slopes of ECAP growth functions from only Hybrid RE CI recipients were significantly correlated with speech performance. The slopes of ECAP growth function in CI24RE users with long electrode also were significantly correlated with performance. However, comparing between 2 independent correlations in RE devices, correlation was higher in Hybrid RE group. CONCLUSION: The results presented in this article support the view that slope of the ECAP growth can show significant correlation to performance with a cochlear implant. Furthermore, these results suggest that the strength of the correlation may be related to the specific device. These results suggest that ECAP measures may be useful in developing a test to predict outcomes with the implant.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。