Inter-device agreement between spectral domain optical coherence tomography, ultrasound biomicroscopy, and gonioscopy in evaluating the iridocorneal angle in normotensive dogs

在评估正常血压犬的虹膜角膜角时,光谱域光学相干断层扫描、超声生物显微镜和房角镜检查之间的设备间一致性

阅读:1

Abstract

BACKGROUND: There has not been a thoroughly reported study of the comparison between spectral domain-optical coherence tomography (SD-OCT) with both ultrasound biomicroscopy (UBM) and gonioscopy on the evaluation of the iridocorneal angle (ICA) in dogs. OBJECTIVES: To investigate the diagnostic value of SD-OCT for the early detection of narrowing ICA by comparing and assessing inter-device agreement in anterior chamber angle (ACA) measurements obtained by SD-OCT and UBM, and ICA evaluations by gonioscopy. METHODS: A total of 28 eyes from 28 client-owned dogs with normal intraocular pressure were included for examination. The ACA and angle opening distance (AOD) were measured from the SD-OCT and UBM images, and gonioscopy images were analyzed using the ICA grade and ZibWest angle index. RESULTS: The mean ACA and AOD for SD-OCT were 28.31° ± 5.37° and 658.42 ± 219.90 µm, and for UBM, 28.34° ± 5.82° and 859.29 ± 221.80 µm, respectively. The mean difference in ACA between the average values of SD-OCT and UBM measurements was 0.03° with a 95% limit of agreement (LoA) span of 16.2°, indicating positive agreement; that in AOD was 200.85 µm with a 95% LoA span of 1,110.95 µm, indicating poor agreement. The Pearson correlation coefficient of the ACA of SD-OCT and ZibWest indices of gonioscopy was 0.624, indicating strong agreement; that of UBM and gonioscopy was 0.43, indicating moderate agreement. CONCLUSIONS: SD-OCT is well tolerated by canine patients due to its non-contact method and might be an alternative option for early screening of ICA narrowing in clinical settings.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。