Airway Scope versus Macintosh laryngoscope: a manikin study

气道镜与麦金托什喉镜:一项人体模型研究

阅读:1

Abstract

OBJECTIVE: To compare the performance of the Airway Scope for tracheal intubation with that of the Macintosh laryngoscope for direct laryngoscopy. METHODS: 10 anaesthetists, 10 residents, 10 nurses and 10 students intubated the trachea of a manikin using the Airway Scope or the Macintosh laryngoscope in each of 4 laryngoscopy scenarios. RESULTS: Within each group of operators, there was no significant difference in the time taken for intubation between the Airway Scope and the Macintosh laryngoscope. However, the overall number of teeth clicks was less in the Airway Scope, and the Airway Scope scored better on the visual analogue scale than did the Macintosh laryngoscope. CONCLUSION: The Airway Scope performed as well as and perhaps better than the Macintosh laryngoscope when used on a manikin.