Transcatheter device closure of atrial septal defects guided completely by transthoracic echocardiography: A single cardiac center experience with 152 cases

经胸超声心动图引导下经导管封堵房间隔缺损:一家心脏中心152例病例的经验

阅读:1

Abstract

OBJECTIVE: This study aimed to assess the safety and feasibility of transcatheter device closure of atrial septal defects (ASDs) guided completely by transthoracic echocardiography (TTE). METHODS: A total of 152 patients underwent transcatheter device closure of ASDs guided completely by TTE in our center from September 2014 to June 2017. We used routine delivery sheaths during the procedure and then closed the ASDs by releasing a domestic occluder. RESULTS: The closure was successful in 150 patients, and surgical repair was required in two patients. The size of the deployed occluder ranged from 10 mm to 38 mm (21.4±8.5 mm), and the procedure duration ranged from 30 to 90 min (38.2±21.4 min). No fatal complications were observed. Minor complications included transient arrhythmias (n=12) during the process of device deployment. The follow-up period was 3 months to 2 years, with no occluder dislodgment, residual fistula, or thrombus-related complications. In our comparative studies, no statistically significant differences were observed in success rates and complications. CONCLUSION: Transcatheter device closure of ASDs guided completely by TTE may be safe and effective and can be an alternative to traditional methods.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。